Capillary Electrophoresis witHEnd-column Amperometric Detection of Urinary 8-Hydroxydeoxyguanosine of Cancer Patients

MEI Su-Rong1,2, CAI Ling-Shuang1, YAO Qing-Hong2, XING Jun1, XU Guo-Wang2, WU Cai-Ying1   

  1. 1. College of Chemistry and Molecular Science, Wuhan University, Wuhan 430072, China;
    2. Analytic Center of State Chromatography Research, Dalian Institute of Chemical Physics, Chinese Academy of Sciences, Dalian 116011, China
  • Received:2002-11-27 Online:2003-11-24 Published:2003-11-24

Abstract: Using one step solid-phase extraction for urine pretreatment and a new sample focusing mode-dynamic pHjunction injection for increasing sensitivity, a simple and sensitive method for the analysis of urinary 8OHdG by capillary electrophoresis witHEnd-column amprometric detection has been developed.The limit of detection was 20 nmol/L(signal to noise ratio S/N=3).The urinary concentration of 8OHdG in nine healthy persons and 28 cancer patients was determined.It was found that urinary concentration of 8OHdG in cancer patients was significantly higher than that in healthy persons[(35.26±27.96) nmol/L) vs.(13.51±5.08) nmol/L, P<0.05], the result demonstrated that 8OHdG may be a good biomarker for measurement of oxidative DNA damage in cancer patients.Furthermore, the excretion levels of urinary 8OHdG from cancer patients receiving surgical therapy are investigated, and the results demonstrated that the smoking has a strong effect on 8OHdG content.

Key words: 8-Hydroxydroxydeoxyguanosine, Capillary electrophoresis witHEnd-column amprometric detection method, Dynamic pH junction injection, Cancer