Improvement of Signal to Noise Ratio in MALDI-TOF MS with Addition of Salts to the Matrix

FANG Cai-Yun1, ZHAI Jian-Jun1, LIU Feng1,2, YANG Peng-Yuan1   

  1. 1. Chemistry Department, Fudan University, Shanghai 200433, China;
    2. Chinese National Human Genome Center at Shanghai, Shanghai 201203, China
  • Received:2004-05-25 Online:2005-04-10 Published:2005-04-10

关键词: 盐, 基质加合物, 信噪比, MALDI-TOFMS

Abstract: CHCA as a matrix facilitates the ionization of proteins and peptides in a matrix-assisted laser desorption/ionization(MALDI) time-of-flight(TOF) mass spectrometer. The matrix itself also ionizes and so do its adducts. Matrix clusters and adducts interfere with peptide ionization and peptide mass spectrum interpretation, especially at low sample concentrations and within m/z 500-1000 range. These matrix adducts, especially within m/z 500-1000, are significantly reduced with addition of ammonium citrate, ammonium sulfate, ammonium benzoate, ammonium bicarbonate, ammonium citrate and potassium dihydrogen phosphate to the matrix and the samples were deposited onto the MALDI target. The reduction of matrix adducts improves peptide ionization and signal-to-noise ratio, which enhances the reliability of database searching.

Key words: Salt, Matrix adduct, S/N ratio, MALDI-TOF MS