Preliminary Research on Pulsed-Electrospray Phenomenon

LU Yu1, BIAN Li-Ping1, ZHOU Feng1, SUN Ping1, YANG Peng-Yuan1, GUO Yin-Long2, YU Chong-Tian2   

  1. 1. Department of Chemistry, Fudan University, Shanghai 200433, China;
    2. Shanghai Institute of Organic Chemistry, Chinese Academy of Sciences, Shanghai 200032, China
  • Received:2000-10-17 Online:2001-04-24 Published:2001-04-24

关键词: 脉冲, 电喷雾, 质谱

Abstract: A novel pulsed electrospray device is reported in this paper. With such an apparatus, the observation of the electrospray process under the pulsed high voltage conditions has been successfully made with microscope, oscilloscope and mass spectrometer. By using pulsed spray technique, the sample flow rate can be averagely decreased to <100 nL/min. After increasing the frequency of pulsed high voltage, it is hopefully to directly combine this device to time of flight mass spectrometer and obtain a great enhancement both in signal to noise ratio and in detection sensitivity for mass spectrometry analysis.

Key words: Pulse, Electrospray, Mass spectrometry