Chem. J. Chinese Universities ›› 1999, Vol. 20 ›› Issue (5): 713.

• Articles • Previous Articles     Next Articles

Computerized Scanning Voltammetric Detector for Capillary Electrophoresis

WANG Li-Shi, YANG Xiao-Yun, MO Jin-Yuan, ZOU Xiao-Yong   

  1. Department of Chemistry, Zhongshan University, Guangzhou, 510275
  • Online:1999-05-24 Published:1999-05-24

Abstract: This work describes the development of a scanning voltammetric detector for capillary electrophoresis. The scanning electrode potential waveform generation and the current signal acquisition are controlled by a microcomputer interfaced with a home-made card which is equipped with 12 bit digital-analog converter(D/A) and a 12 bit analog-digital converter (A/D). The D/A is connected to a Ag/AgCl reference electrode and used to control the cell potential via a voltammeter, and the A/D is connected to the working electrode for data acquisition. The methods of potential control and signal acquisition are discussed. An example of separating mixture of p-diphenol and pyrocatechol is demonstrated. The results obtained are satisfactory.

Key words: Capillary electrophoresis, Microcomputer controlling, Scanning voltammetric detector

CLC Number: 