Chem. J. Chinese Universities ›› 1989, Vol. 10 ›› Issue (1): 68.

• Articles • Previous Articles     Next Articles

Studies on Clonidine Hydrochloride Selective Coated Wire Electrode with Bilayer Polymeric Films

Hu Xiaoya, Leng Zongzhou   

  1. Department of Chemistry, Yangzhou Teachers College, Yangzhou
  • Received:1987-07-20 Online:1989-01-24 Published:1989-01-24

Abstract: A new type of coated wire electrode (CWE) with bilayer .polymeric films is reported. Platinum wire is modified by conductive polyaniline and then with a film of PVC (containing electroactive species). This kind of electrode not only has merits of CWE, but also has fine stability and repro-ducibility. With successive determination of 8 hours, the potential drifts are about 0.8mV. The Nernstian linear range of electrode shows 1 × 10-1~8×10-6mol/L with an average slope of 56.7mV/decade. The direct potentiometry for samples shows 100.9% of average recovery and 1.3% of standard deviation.

Key words: Potentiometry, Ion-selective electrode, Coated-wire electrode, Modified electrode, Clonidine