The Determination of Anti-MV Antibody in Serum with CL-ELISA

ZHANG Zhu-Jun1, ZOU Ke-Wei1, CHENG Ming-Jie2   

  1. 1. Institute of Analytical science, Shaanxi Normal University, Xian, 710062;
    2. Shaaanxi Sanitation and Antiepidemic Station
  • Received:1993-05-10 Revised:1993-09-20 Online:1994-04-24 Published:1994-04-24

Abstract: A method for determination of the anti-MVantibody in serum is established based on chemiluminescence enzyme-linked immunosorbant assay (CL-ELISA).The MV antigen is coated on the polystyrene cup, then reacts with anti-MV serum to form the immunocompound.The anti-HuIgGlabeled by HRPis used as a second antibody.The enzymatic activity on the antibody is determined by p-iodophenol enhanced luminol-H2O2 chemiluminescent system, which has higher sensitivity than current chromogenic reaction.The curve of anti-MV antibody title determined with CL-ELISA and ELISA is established and 30 clinical samples are tested.The results show that the CL-ELISAhas higher positive ratio than ELISA.Two methods have good correlation.The correlation coefficient is 0.992.

Key words: Anti-MV antibody, Chemiluminescence, Enzyme Immunoassay, HRP