Chem. J. Chinese Universities ›› 2006, Vol. 27 ›› Issue (4): 618.

• Articles • Previous Articles     Next Articles

Sample Pretreatment Microfluidic Chip for DNA Extraction from Rat Peripheral Blood

CHEN Xing1, CUI Da-Fu 1*, LIU Chang-Chun1, LI Hui1, ZHAO Wei-Xing2   

  1. 1. State Key Lab of Transducer Tech., Institute of Electronics, Chinese Academy of Sciences, Beijing 100080, China; 2. State
    Key Laboratory of Protein and Plant Genetic Engineering, College of  Life Sciences, Peking University, Beijing 100871, China
  • Received:2005-06-20 Online:2006-04-10 Published:2006-04-10
  • Contact: CUI Da-Fu, dfcui@mail.ie.ac.cn

Abstract:

A sample pretreatment microfluidic chip was described based on the principle of solid phase extraction(SPE) and the MEMS technology. Oxidized porous silicon with a large surface area as the solid phase matrix for absorption DNA from biologic sample can greatly improve the DNA yield. The factors which can affect the DNA yield were analyzed and the preparation technology and the experiment procedure were improved. The DNA purification process from the rat peripheral blood can be achieved and the DNA yield is 24  ng/(μL whole blood), which can reach the level of the commercial DNA purification kits. Furthermore, the DNA extracted from the whole blood can be amplified by PCR, which can achieve a high efficiency of the amplification.

Key words: MEMS; Oxidized porous silicon; Sample pretreatment microfluidic chip

CLC Number: 