Chem. J. Chinese Universities ›› 2001, Vol. 22 ›› Issue (8): 1301.

• Articles • Previous Articles     Next Articles

Determination of Cadmium at Ultra-trace Level by FI Cold Vapor Generation AAS with On-line Ion Exchange Preconcentration

YAO Bo, XU Shu-Kun, FANG Zhao-Lun   

  1. Research Center for Analytical Sciences, Northeastern University, Shenyang 110006, China
  • Received:2000-05-17 Online:2001-08-24 Published:2001-08-24

Abstract: Amethod for the determination of cadmium by flow injection cold vapor generation atomic absorption spectrometry(FI-CVGAAS) with an on line ion exchange micro column preconcentration system was developed using IRC718, 122 and Muromac A-1 resin. The interferences from coexisting elements were studied. The detection limit(3σ) was 2.3-3.0 ng/Lwith a sampling frequency of 45-60/h. The precision was 2.0%-2.6% RSD( n=11) at 0.2 μg/Llevel. The method was applied to the analysis of water samples. Satisfactory recoveries of cadmium in reservoir water, tap water and brine waters were obtained for spiked samples.

Key words: Cadmium, Flow injection(FI), Ion-exchange preconcentration, Cold vapor generation, Atomic absorption spectrometry