Chem. J. Chinese Universities ›› 2000, Vol. 21 ›› Issue (11): 1664.

• Articles • Previous Articles     Next Articles

Immobilization of Antigen and Antibody on Gold Surface and Determination Using Fluorescent Method

JIANG Xiong-Ping2, XU Dan-Ke1, LIU Yao-Qing1, MA Li-Ren1   

  1. 1. Institute of Radiomedicine, Academy of Military Medical Sciences, Beijing 100850, China;
    2. Institute for Drug Control of PLA, Beijing 100071, China
  • Received:2000-01-21 Online:2000-11-24 Published:2000-11-24

Abstract: A Gold surface was modified with N-acetylcysteine to produce a self-assembled monolayer and then a single layer film of rabbit IgG was immmobilized on modified gold surface covalently using carbodiimide hydrochloride(EDC) and N-hydroxysuccinimide(NHS). The competitive method(Ⅰ), the sandwich method(Ⅱ) and the surplus method(Ⅲ) were developed to determine rabbit IgG or goat anti rabbit IgG, respectively. A luminescence analyzer connected with a fluorescence microscope was employed to detect the light intensity of labeled sheep anti rabbit IgG. The results of sensitivity and dynamic range were obtained as 25 mg/L and 25—600 mg/L for method(Ⅰ), 1.0 mg/L and 1—200 mg/L for method(Ⅱ) and 25 mg/L and 25—500 mg/L for method(Ⅲ), respectively.

Key words: Immunosensor, Self-assembled monolayer, Fluorescence

CLC Number: 

TrendMD: