Chem. J. Chinese Universities ›› 1992, Vol. 13 ›› Issue (9): 1192.

• Articles • Previous Articles     Next Articles

Preparation of Film Modified Electrode by Electrochemical Method and Its Application for Anodic Stripping Voltammetric Determination of Pb(Ⅱ)

ZHUANG Xiu-Run, XU Jin-Rui   

  1. Department of Applied Chemistry, Huaqiao University, Quanzhou, 362011
  • Received:1991-01-28 Revised:1992-04-13 Online:1992-09-24 Published:1992-09-24

Abstract: In this paper we described the glassy carbon electrode treated by an electrochemical method in an aqueous solution containing the o-aminophenol,formaldehyde and sodium hydroxide.The sensitivity of the treated electrode for Pb2+ is about 60 times higher than that of untreated glassy carbon electrode.The electrode has a good reproducibility with a relative standard deviation of 3.7%(n=10).The anodic stripping peak height depends linearly on the Pb2+ concentration up to 10 ng/mL,and the quantitative limit with this electrode is 0.2 ng/mL.

Key words: Lead, Glassy-carbon electrode, Anodic stripping voltammetry, o-Aminophenol

TrendMD: