A Micro-modified Electrode Detecting Nitric Oxide

DING Cai-Rong, Mo Jin-Yuan, CAI Pei-Xiang   

  1. Department of Chemisty, Zhongshan University, Guangzhou 510275, China
  • Received:1998-10-27 Online:1999-09-24 Published:1999-09-24

Abstract: The effects of the adduct of imidazole and [Fe(Sal2en)]2+oxidizing nitric oxide electrocatalytically were studled in this paper.The adduct of the 1: 1 mole ratio of Im/[Fe(Sal2en)]2+can improve twice the sensitivity at which [Fe(Sal2en)]2+oxidizes nitric oxide.It is certain that the media NO·[Fe(Sal2en)]2+generates during the electro-oxidation of nitric oxide.The electrode above has a good linear response in the range of 3×10-6~2×10-9mol/Lof NO.The detection limit is 2×10-9mol/L, RSD%<0.5%, δ>0.9994.It satisfys with the need of the phsiological analysis.

Key words: Nitric oxide, Imidazole, [Fe(Sal2en)]2+, Electrocatalytic-oxidation