Abstract: I-may be oxidized to form IO3-.On this basis The determination of iodine ion by dif-ferential pulse anodic st ripping voltammetry(DPAS)on Pt ball work Electrode was investigated. For overcoming The absorption of IO3-, The Electrode was washed by Electrochemical method(at+1.3 V)for 120s.This method can detect 5.00×10-7mol/L I-in 0.1 mol/L KH2PO42K2HPO4buffer solution(pH=4).Agood proportionality between The peak currentof DPAS and iodineion concent ration in The range of 5.00×10-7~1.50×10-5mol/L has been observed. This met hod was successf ully applied to determination of iodine in kelp and edible salt.

Key words: Differential pulse anodic stripping voltammetry, Iodine, Kelp, Edible salt