Chem. J. Chinese Universities ›› 1989, Vol. 10 ›› Issue (12): 1211.

• Articles • Previous Articles     Next Articles

The Construction and Calibration of a Precision Solution Calorimeter

Zou Lizhuang1, Zhang Zhonghai2, Xu Guiduan3, Qu Songsheng3   

  1. 1. Shandong Liaocheng Teacher's College, Liaocheng;
    2. Hubei Xiaogan Teacher's Speciality School, Xiaogan;
    3. Department of Chemistry, Wuhan University, Wuhan
  • Received:1988-03-01 Online:1989-12-24 Published:1989-12-24

Abstract: Aprecision solution calorimeter was constructed in the laboratory.Athermsistor thermometer was used for temperature measurement.The calorimeter was calibrated with calorimetric standardreference materials——spectroscopically pure potassium chloride and tris (hydroxymethyl)aminomethane from N.B.S.-724 a in USA.The results obtained at 25℃ was found to be 17518± 15 J.mol-1 for the enthalpy of solution of potassium chloride in water and -29729 ±31 J.mol-1 for the enthalpy of reaction of TRISin 0.1008 mol/L HCl.The values are well in agreement with the published results.The reliability of this apparatus was confirmed

Key words: Tetraphenylborates, Thermochemistry, Heat of solution, Calorimeter, Calibration