The Enhance of the Sensitivity of Conventional Flame Atomic-Absorption Spectrometry by Slotted Stainless Steel--Determination of Copper and Zinc in Serum

Sun Shuju, Bing Guide, Zheng Yansheng, Guo Yi   

  1. Department of Chemistry, Jilin University, Changchun
  • Received:1986-08-18 Online:1987-11-24 Published:1987-11-24

Abstract: A slotted stainless steel tube when used in air-acetylene flame increses the residence time of atoms in the light path of an atomic-absorption spectrometer, thus increasing the analytical sensitivity. With this technique copper and zinc were determined in serum diluted 1:20. Rapid accurate analyses can be achieved by using continuous aspiration. Use of aqueous standards resulted in recovery for copper and zinc of 92-110.9%. Coefficient of variation of 2.7% and 3.6% respectively.