高等学校化学学报 ›› 1992, Vol. 13 ›› Issue (5): 599.

• 论文 • 上一篇    下一篇

IAP-4S螯合物和离子对反相高效液相色谱测定铁、钴和镍

林长山1, 张孝松1, 刘学著1, 冯建章2, 童沈阳2   

  1. 1. 中国科学技术大学应用化学系, 合肥, 230026;
    2. 北京大学化学系
  • 收稿日期:1991-04-12 出版日期:1992-05-24 发布日期:1992-05-24
  • 通讯作者: 林长山

Determination of Fe, Co and Ni in Their Chelates with IAP-4S by Ion-Pair Reversed-Phase High-Performance Liquid Chromatography

LIN Chang-shan1, ZHANG Xiao-song1, LIU Xue-zhu1, FENG Jian-zhang2, TONG Shen-yang2   

  1. 1. Dept.of Applied Chem., University of Science and Technology of China, Hefei, 230026;
    2. Department of Chemistry, Peking University, Beijing
  • Received:1991-04-12 Online:1992-05-24 Published:1992-05-24

关键词: 铁、钴和镍的测定, 2-(2-咪唑偶氮)-苯酚-4-磺酸(IPA-4S), 镍矿

Abstract: The chelates of iron, cobalt and nickel with 2-(2-imidazolylazo)-phenol-4-sulphonic acid IAP-4Swere separated on a column (250×4 mm) of Nucleosil C18(5μm) with a mobile phase (0.7 mL/min) of methanol-water (30:70, V/V) containing 0.01 mol/L KH2PO4-Na2HPO4 buffer (pH 7.5) and 2×10-3 mol/Ltetrabutyl ammonium bromide, and were detected at 550 nm.The calibration graphs are rectilinear from 0.1~5 ppm for iron, 0.05~5 ppm for both cobalt and nickel.The detection limits are 0.25 ng for iron, 0.16 ng for both cobalt and nickel.The method was applied to the analysis of real samples.

Key words: Determination of iron, cobalt and nickel, 2-(2-imidazolylazo)-phenol-4-suIphonic acid, Nickel ore

TrendMD: