Chem. J. Chinese Universities ›› 1995, Vol. 16 ›› Issue (12): 1871.

• Articles • Previous Articles     Next Articles

The Application of On-line Standard Addition Method in Microwave Plasma Torch Atomic Emission Spectrometry

YE Dong-Mei, ZHANG Han-Qi, YU Jing-Lei, Jin Qin-Han   

  1. Department of Chemistry, Jilin University, Changchun, 130023
  • Received:1994-12-20 Revised:1995-08-20 Online:1995-12-24 Published:1995-12-24

Abstract: In this paper, the probability of eliminating matrix interferences in MPT-AES by on-line standard addition method was studied. The dynamic linear range of Cu and Mg is 0. 05~80 μg/mL.The relative standard deviations of them are 3.02% (1.0 μg/mL, n=7).and 4.47 % (1. 0 μg/mL, n=7) respectively. The results indicated that this method can eliminate matrix interferences and is useful in real sample analysis. The operation of this system is simple.

Key words: On-line standard addition method, Microwave plasma torch, Atomic emission spectrometry

TrendMD: