Modification on the Interface of Capillary Isoelectric Focusing and Mass Spectrometry for Proteomic Research

XU Chong-Feng1, SONG Hao-Wei1, YANG Peng-Yuan1, WANG Qian1, WANG Hong-Hai2, XU Yun-Min2   

  1. 1. Department of Chemistry, Shanghai 200433, China;
    2. Institute of Genetics, Fudan University, Shanghai 200433, China
  • Received:2001-11-06 Online:2002-06-24 Published:2002-06-24

Abstract: An important interface modification for capillary isoelectric focusing mass spectrometry(CIEF-MS) is reported. Acapillary needle with a sheath flow tube pre-mounted in one device serves as both the CIEF cathodic-end and the ESI interface. The needle was inserted into the cathode reservoir at the focusing stage, and then removed to the front of sampler orifice of MSat the detection stage. Therefore, the shut-off of the high voltage is not necessary and a better focusing protein zone is maintained during the ESIstage. The optimization of the experimental condition has been investigated and the two dimensional image of the E. coli proteins has been obtained, which is comparable with that from a 2D-gel.

Key words: Interface of CIEF-MS, 2D Separation and identification, E. coli, Proteome