Chem. J. Chinese Universities ›› 2000, Vol. 21 ›› Issue (5): 694.

• Articles • Previous Articles     Next Articles

Direct Determination of Trace Impurity Yttrium in Titantium Oxide Powder by Fluorination Assisted ETV-ICP-AES Using Slurry Sampling

PENG Tian-You, YAN Qin, HU Bin, JIANG Zu-Cheng   

  1. Department of Chemistry, Wuhan University, Wuhan 430072, China
  • Received:1999-07-21 Online:2000-05-24 Published:2000-05-24

Abstract: By using a polytetrafluorethylene(PTFE) emulsion as fluorinating reagent, a method for fluorinating assisted eletrothermal vaporization (ETV)/ICP-AESdirect determination of trace yttrium in TiO2 ceramic powder has been proposed; the various facts which would affect the vaporization process between the matrix and the analyte were studied ; the fluorinating vaporization behavior of the matrix and the analyte both in slurry and in solution were investigated. Based on pre-separation between the matrix and the analyte, the matrix effect was reduced. On the optimum experiment conditions, the detection limit of the present method is 0.26μg/L; and the RSDis 3.8 % (n=5, c=0.5 mg/L). This method has been applied to directly determinate trace yttrium in two kinds of TiO2 powder without any chemical pretreatment.

Key words: Fluorination electrothermal vaporization, Slurry sampling, ICP-AES, Trace impurity yttrium

CLC Number: 

TrendMD: