Abstract:
A method was presented for the determination by Inductively Coupled Plasma Mass Spectrometry (ICP-MS) of Impurity Cerium in Potassium Lithium Niobate (K3Li2Nb5O15) Crystal. A standard model Perkin-Elmer/SCIEX ELAN 5000 ICP Mass Spectrometer was used. The instrument was operated at a RF power setting of 1.05 kW. A standard demountable quartz torch with 1.2mm id alumina injector tube was used with the following argon gas flow rates:nebulizer 0.84L/min,auxiliary 0.80L/min,and coolant 15.0L/min.