Investigation on Triangular-wave AC Oscillopolarography(Ⅰ)——Properties of Triangular-wave AC Oscillopolarogram

Qi Hong, Ma Xinsheng, Gao Hong(H. Kao)   

  1. Department of Chemistry, Nanjing University, Nanjing
  • Received:1986-12-30 Online:1988-06-24 Published:1988-06-24

Abstract: Triangular-wave current sweeping was used in ACoscillopolarography instead of sine-wave current. The oscillopolarogram obtained is similar to that obtained with sine-wave current. The appearance or disappearance of incision on the oscillopolarogram can be used to indicate the end-point of titration. The relationship of incision height hi to the concentration of depolarizer and current sweeping rate Vare respectively as follows.hi(mV) = -0.02300 + 4.702×10-4C0*(-2/3) hi(mV)= -0.02807 +950.68V(2/3) When Vis very lower 5.0×10-9mol/L Tl+ ion can be determined by this method

Key words: AC Oscillopolarography, Triangular-wave current sweeping