Chem. J. Chinese Universities ›› 1985, Vol. 6 ›› Issue (8): 689.

• Preface • Previous Articles     Next Articles

Direct Determination of Trace Amount of Iron in the Phosphate by the Polaried Zeeman A. A. S.

He Jinlan, Tang Yurong   

  1. Test Centre, Wuhan University, Wuhan
  • Received:1984-06-23 Online:1985-08-24 Published:1985-08-24

Abstract: In this paper, the determination of the trace amount of iron in the phosphate by AASwith graphite furnace atomizer is discussed. It was found that the interference of PO4-3 and some metal for iron can be suppressed by the use of the higher charring temperature (1300℃) and Zeeman effect.After the sample is dissolved, it can be measured directly, without using any way of separation.

TrendMD: