高等学校化学学报 ›› 2000, Vol. 21 ›› Issue (11): 1661.
• 研究简报 • 上一篇 下一篇
丁兰1, 金海燕1, 刘淼2, 梁枫2, 郇延富2, 曹彦波2, 张寒琦2, 金钦汉2
收稿日期:
出版日期:
发布日期:
通讯作者:
基金资助:
国家自然科学基金(批号:29735160)资助
DING Lan1, JIN Hai-Yan1, LIU Miao2, LIANG Feng2, HUAN Yan-Fu2, CAO Yan-Bo2, ZHANG Han-Qi2, JIN Qin-Han2
Received:
Online:
Published:
关键词: 气动雾化器, 热雾化器, 微波热雾化器, ICP-AES, 进样方法
Abstract: The analytical performance of two sample introduction systems, a low-microwave powered thermospray nebulizer and a pneumatic nebulizer, coupled to an inductively coupled plasma atomic emission spectrometer, was compared. Several parameters including the sample uptake rate, the carrier gas flow rate and the observation height(above the load coil) were optimized and compared. Under the optimized conditions, the background equivalent concentration of Mn 257 610 nm, memory effect, HCl concentration effect on emission intensity of some selected elements, detection limits(DLs) and relative standard deviation(RSD) for Cd, Co, Cr, Mn, La with either nebulizer were studied. The DLs obtained with MWTN are lower than those with PN, but the RSDs are worse with MWTN than with PN.
Key words: Pneumatic nebulizer, Thermospray, Microwave-powered thermospray, ICP-AES, Sample introduction
中图分类号:
O657
TrendMD:
丁兰, 金海燕, 刘淼, 梁枫, 郇延富, 曹彦波, 张寒琦, 金钦汉. 用低功率微波热雾化器与十字交叉气动雾化器作为ICP-AES进样装置的比较. 高等学校化学学报, 2000, 21(11): 1661.
DING Lan, JIN Hai-Yan, LIU Miao, LIANG Feng, HUAN Yan-Fu, CAO Yan-Bo, ZHANG Han-Qi, JIN Qin-Han. Comparison of a Low Microwave-powered Thermospray Nebulizer and a Pneumatic Nebulizer for ICP-AES. Chem. J. Chinese Universities, 2000, 21(11): 1661.
0 / / 推荐
导出引用管理器 EndNote|Ris|BibTeX
链接本文: http://www.cjcu.jlu.edu.cn/CN/
http://www.cjcu.jlu.edu.cn/CN/Y2000/V21/I11/1661