Chem. J. Chinese Universities ›› 1989, Vol. 10 ›› Issue (7): 709.

• Articles • Previous Articles     Next Articles

Study on the Technique of ICP-AES by Using Electrothermal Graphite Furnace as Sample Introduction Device——Device and Its Analytical Performance

Huang Min, Xu Lianfa, Jiang Zucheng, Zeng Yun'e   

  1. Department of Chemistry, Wuhan University, Wuhan
  • Received:1987-12-23 Online:1989-07-24 Published:1989-07-24

Abstract: An instrument system is described in which a graphite furnace electrothermal vaporization device is used for the.introduction of microlitre liquid sample into an inductively coupled plasma.The appratus is equipped with home-made ones.Under the optimum operating condition for ICPexcitation and graphite furnace evaporation, the detection limits are at sub-nanogram level with linear dynamic ranges of over four orders of magnitude for nine elements.The RSDare better than 5%.

Key words: Graphite furnace, Electrothermal vaporization, Inductively coupled plasma(ICP), Combined technique