全二维液相色谱(NPLC×RPLC)接口及其应用

田宏哲, 徐静, 关亚风   

  1. 中国科学院大连化学物理研究所仪器分析研究室, 大连 116023
  • 收稿日期:2006-11-21 修回日期:1900-01-01 出版日期:2007-04-10 发布日期:2007-04-10
  • 通讯作者: 关亚风

Sample Loops-valve Interface of Comprehensive Two-dimensional Liquid Chromatography(NPLC×RPLC) and Its Application

TIAN Hong-Zhe, XU Jing, GUAN Ya-Feng*   

  1. Department of Instrumentation & Analytical Chemistry, Dalian Institute of Chemical Physics, Chinese Academy of Sciences, Dalian 116023, China
  • Received:2006-11-21 Revised:1900-01-01 Online:2007-04-10 Published:2007-04-10
  • Contact: GUAN Ya-Feng

摘要: 用内径为0.53 mm的填充毛细管正相液相色谱为第一维, 用4.6 mm(i.d.)×50 mm RP-18e整体柱反相色谱为第二维, 建立了定量环-阀切换接口的全二维液相色谱系统(NPLC×RPLC). 第一维色谱分离洗脱出的组分交替存储在十通阀上的两个定量环中, 同时定量环中前一个组分被转移到第二维进行反相分离. 因为第一维的流动相流量仅是第二维的1/500, 自然解决了流动相兼容问题. 采用芳香族化合物的混合物和中药丹参正己烷提取液对该全二维液相系统的分离能力进行了评价.

关键词: 全二维液相色谱, 正相色谱/反相色谱联用, 定量环-阀切换接口, 中药, 丹参

Abstract: A comprehensive two-dimensional liquid chromatography(NPLC×RPLC) was developed. Normal phase liquid chromatography(NPLC) with a CN microcolumn was used as the first dimension, and reversed-phase liquid chromatography(RPLC) with a monolithic column was used as the second dimension. The interface was based on a ten-port, two-position valve with two storage loops. The first column was operated at a flow rate of 8 μL/min in isocratic mode, while the monolithic column flow rate was 4 mL/min and was operated in gradient mode. The sample loops had a volume of 12 μL each, and the analysis time in the second dimension was 1.5 min. The primary fractions were stored into the two storage loops alternately, and then were separated by the secondary column sequentially. AUV detector was used after the secondary column. The use of a monolithic column in the second dimension permits a very fast analysis operating at highter flow rates without loss of resolution, since the monolithic column has higher permeability than conventional packed columns. The incompatibility of the two dimensional mobile phases was avoided because the flow rate of the first dimension is only 1/500 of that of the second dimension. The performance of the system was evaluated by several standards mixture and Radix salviae miltiorrhiza bage extract and the advantage of using 2D-LC over one-dimensional chromatographic techniques is clearly demonstrated in the analysis of the extract.

Key words: Comprehensive two-dimensional LC, NPLC×RPLC coupling technique, Sample loops-valve interface, Traditional Chinese medicine, Radix salviae miltiorrhiza bage